Standard data logger for recording pressure and temperature values. Easy to program and install, no data loss during validation process, multiple users can share the same database simultaneously. Factory calibration on users site.
• Collects and stores temperature information in the internal memory
• Recover the data through the docking station using XpertLog® Software
• ITS-90 calibration coefficients are stored in the internal memory
• Complies with FDA 21 CFR Part 11
• Network ready: multiple users, centralized, server database, active directory authentification
• Calibration can be performed on site
±10 mbar ≤ 3 BarA @ 130°C
±15 mbar > 3 BarA @ 130°C
±15 mbar ≤ 3 BarA @ 23°C
±30 mbar > 3 BarA @ 23°C
±0.1°C from -40°C to +140°C
Resolution
0.1 mbar; 0.01°C
Memory capacity
16,000 data points pressure / 16,000 data points temperature
Memory type
non volatile
Sampling rate
1 sec to 24 hours
Internal clock drift
4 sec / 24 h @ 23°C
Mechanical
- material
Peek 1000 - FDA Compliant
- enclosure
IP-51
- weight
57 g (2.01 oz)
- diameter
40 mm (1.57")
- height
69 mm (2.72")
Power
1 x Lithium batteries, 0.35 g/each Lithium content
classified UN3091 under IATA DGR section 3.9.2.6
compliant with ICAO/IATA packing instructions 970,
user replaceable
Battery lifetime
Up to 12 months, based on usage
Docking Station
USB to computer
4, 8, 12 and 20 ports available
Programs and reads unlimited number of loggers
Calibrations
Factory Calibrations:
User Calibrations:
traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory
closed loop calibration using XpertLog® Software
Standard Calibration Points
hysteresis at 23°C and 130°C:
• 0.05 BarA
• 0.50 BarA
• 1.00 BarA
• 1.50 BarA
• 2.00 BarA
• 2.50 BarA
• 3.00 BarA
• 3.50 BarA
• 4.00 BarA
• 4.50 BarA
• 5.00 BarA