Real time wireless temperature and CO2 data logger with flexible sensor
Wireless data logger GMP RF TCO2 F
0% to 20% CO2 0°C to +50°C 300 mBar to 1100 mBar (built-in pressure compensation)
GMPF CO2 RF collects real time CO2, temperature and pressure values and sends the data to XpertLog® Software. With an accuracy of ±0.5% CO2 and ±0.1°C, GMPF CO2 RF allows you to measure CO2 up to 20% and temperature from -200°C to +200°C.
• Collects and transmits real time temperature to XpertLog Software • ITS-90 Calibration coefficients are stored in the internal memory • Complies with FDA 21 CFR Part 11 • Network ready: multiple users, centralized, server database, active directory authentification • Calibration can be performed on site • Can be remotely activated
0% to 20% CO2 0°C to +50°C 300 mBar to 1100 mBar (built-in pressure compensation)
- sensor
-200° to +200°C
Accuracy
±0.1°C from -90°C to +140°C ±0.5% CO2 from 0% to 10% CO2
Resolution
0.01% CO2 / 0.01°C
Memory capacity
16,000 data points CO2 / 16,000 data points temperature
Memory type
non volatile
Sampling rate
1 sec to 24 hours
Transmission
simultaneous: maximum 24 loggers sequential: unlimited
Internal clock drift
4 sec / 24 h @ 23°C
Mechanical
- material
Peek 1000 - FDA Compliant
- enclosure
IP-30
- weight
115g153 g (5.39 oz) (4.06 oz)
- diameter
65 mm (2.56")
- height
52 mm (2.04")
Power
2 x Lithium batteries, 0.35 g/each Lithium content classified UN3091 under IATA DGR section 3.9.2.6 compliant with ICAO/IATA packing instructions 970, user replaceable
Battery lifetime
Up to 6 months, based on usage
Antenna
Internal
Wireless range
121 m (400 feet)
Radio frequency
2.4 GHz ISM band
Access point
USB to computer Programs and reads unlimited number of loggers
Calibrations
Factory Calibrations: traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory User Calibrations: closed loop calibration using XpertLog® Software
Standard Calibration Points
0°C +23°C +50°C
Certifications
EMC: EN 60601-1-2 / 2007 RF: EN 300440-2 v.1.3.1 (2009-03)