Real Time Wireless Temperature Data Logger with Flexible Sensor
Wireless Data Logger GMP RF Bio
-90°C to +140°C
Real time temperature data logger created for microbial detection systems. High accuracy (±0.1°C), quick validation time (no wires or pre-calibration required).
+ Collects and transmits real time temperature to XpertLog Software
+ ITS-90 Calibration coefficients are stored in the internal memory
+ Complies with FDA 21 CFR Part 11
+ Network ready: multiple users, centralized, server database, active directory authentification
+ Calibration can be performed on site
+ Can be remotely activated
Order Reference

GMP-RF-Bio-0.5m

Specifications
Characteristics
LVSC.3.2.0 - January 2022
Sensor Type
4 wires PT-100
- flexible
430 mm (17")
- customized
up to 5000 mm (195")
- sensor size
1.5 x 8 mm (0.06 x 0.315")
- cross section
1.77 mm2 (0.0027 inch2)
- body
-90°C to +140°C
- sensor
-200° to +200°C
Accuracy
±0.1°C from -90°C to +140°C
Resolution
0,01°C
Memory capacity
32,000 data points
Memory Type
non volatile
Sampling rate
1 sec to 24 hours
Transmission
simultaneous: maximum 20 loggers
sequential: unlimited
Internal clock drift
4 sec / 24 h @ 23°C
Mechanical
- material
Peek 1000 - FDA Compliant
- enclosure
IP-68
- weight
67 g (2.36 oz)
- diameter
33.5 mm (1.32")
- height
101.5 mm (4.00")
Power
2 x Lithium batteries, 0.35 g/each Lithium content
classified UN3091 under IATA DGR section 3.9.2.6
compliant with ICAO/IATA packing instructions 970,
user replaceable
Battery lifetime
Up to 6 months, based on usage
Antenna
Internal
Wireless range
121 m (400 feet)
Radio frequency
2.4 Ghz ISM band
Access point
USB to computer
Programs and reads unlimited number of loggers
Calibrations
Factory Calibrations:
User Calibrations:
traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory
closed loop calibration using XpertLog® Software
Standard Calibration Points
-80°C
0°C
+140°C
Certifications
EMC:
Health Safety:
RF:
EN 60601-1-2 / 2007
EN 60950-1 / 2006
EN 300440-2 v.1.3.1 (2009-03)