Quick navigation

Real Time Wireless Temperature and CO2 Data Logger with Flexible Sensor

Wireless Data Logger GMP RF TCO2 F

0% to 20% CO2
0°C to +50°C
300 mBar to 1100 mBar (built-in pressure compensation)

GMPF CO2 RF collects real time CO2, temperature and pressure values and sends the data to XpertLog® Software. With an accuracy of ±0.5% CO2 and ±0.1°C, GMPF CO2 RF allows you to measure CO2 up to 20% and temperature from -200°C to +200°C.

+ Collects and transmits real time temperature to XpertLog Software
+ ITS-90 Calibration coefficients are stored in the internal memory
+ Complies with FDA 21 CFR Part 11
+ Network ready: multiple users, centralized, server database, active directory authentification
+ Calibration can be performed on site
+ Can be remotely activated

Order Reference

GMP-RF-TCO2-F

Specifications

Characteristics

LVSC.3.2.0 - January 2022

Sensor Type

CO2 - IR / Temp: 4 wires PT-100

- flexible

430 mm (17")

- customized

up t0 5000 mm (195")

- sensor size

1.5 x 8 mm (0.06 x 0.315")

- cross section

1.77 mm2 (0.0027 inch2)

- body

0% to 20% CO2
0°C to +50°C
300 mBar to 1100 mBar (built-in pressure compensation)

- sensor

-200°C to +200°C

Accuracy

±0.1°C from -90°C to +140°C
±0.5% CO2 from 0% to 10% CO2

Resolution

0.01% CO2 / 0.01°C

Memory capacity

16,000 data points CO2 / 16,000 data points temperature

Memory type

non volatile

Sampling rate

1 sec to 24 hours

Transmission

simultaneous: maximum 20 loggers
sequential: unlimited

Internal clock drift

4 sec / 24 h @ 23°C

Mechanical

- material

Peek 1000 - FDA Compliant

- enclosure

IP-30

- weight

153 g (5.39 oz)

- diameter

64.5 (2.54")

- height

52 mm (2.04")

Power

2 x Lithium batteries, 0.35 g/each Lithium content
classified UN3091 under IATA DGR section 3.9.2.6
compliant with ICAO/IATA packing instructions 970,
user replaceable

Battery lifetime

Up to 6 months, based on usage

Antenna

Internal

Wireless range

121 m (400 feet)

Radio frequency

2.4 Ghz ISM band

Access point

USB to computer
Programs and reads unlimited number of loggers

Calibrations

Factory Calibrations:
User Calibrations:

traceable NIST/COFRAC - ITS-90 coefficients stored in the internal memory
closed loop calibration using XpertLog® Software

Standard Calibration Points

• 0°C
• 23°C
• 50°C

Certifications

EMC:
Health Safety:
RF:

EN 60601-1-2 / 2007
EN 60950-1 / 2006
EN 300440-2 v.1.3.1 (2009-03)

Find out more about XpertVal 64

Fill out this form to obtain a demo / test.

Please enable JavaScript in your browser to complete this form.
Address